In medicine, the pulse is the rhythmic throbbing of each artery in response to the cardiac cycle (heartbeat).[1] The pulse may be palpated in any place that allows an artery to be compressed near the surface of the body, such as at the neck (carotid artery), wrist (radial artery or ulnar artery), at the groin (femoral artery), behind the knee (popliteal artery), near the ankle joint (posterior tibial artery), and on foot (dorsalis pedis artery).[2][1] The pulse is most commonly measured at the wrist or neck.[3] A sphygmograph is an instrument for measuring the pulse.[4]

Pulse
Diagram of the rise and lower of blood from a pulse.
OrganismsAnimalia*
Biological systemCirculatory system
ActionInvoluntary
MethodHeart pumps blood using reciprocating method causing inconstant blood flow throughout the circulatory system that can be recognized. (See Cardiac cycle)
Frequency60–100 per minute (humans)
Duration0.6–1 second (humans)
Animalia with the exception of Porifera, Cnidaria, Ctenophora, Platyhelminthes, Bryozoan, Amphioxus.

Physiology

edit
 
Pulse evaluation at the radial artery.
 
Recommended points to evaluate pulse

Claudius Galen was perhaps the first physiologist to describe the pulse.[5] The pulse is an expedient tactile method of determination of systolic blood pressure to a trained observer. Diastolic blood pressure is non-palpable and unobservable by tactile methods, occurring between heartbeats.

Pressure waves generated by the heart in systole move the arterial walls. Forward movement of blood occurs when the boundaries are pliable and compliant. These properties form enough to create a palpable pressure wave.

Pulse velocity, pulse deficits and much more physiologic data are readily and simplistically visualized by the use of one or more arterial catheters connected to a transducer and oscilloscope. This invasive technique has been commonly used in intensive care since the 1970s.

The pulse may be further indirectly observed under light absorbances of varying wavelengths with assigned and inexpensively reproduced mathematical ratios. Applied capture of variances of light signal from the blood component hemoglobin under oxygenated vs. deoxygenated conditions allows the technology of pulse oximetry.

Characteristics

edit

Rate

edit

The rate of the pulse can be observed and measured on the outside of an artery by tactile or visual means. It is recorded as arterial beats per minute or BPM. Although the pulse and heart beat are related, they are not the same. For example, there is a delay between the onset of the heart beat and the onset of the pulse, known as the pulse transit time, which varies by site. Similarly measurements of heart rate variability and pulse rate variability differ.[6]

In healthy people, the pulse rate is close to the heart rate, as measured by ECG.[6] Measuring the pulse rate is therefore a convenient way to estimate the heart rate.[7] Pulse deficit is a condition in which a person has a difference between their pulse rate and heart rate. It can be observed by simultaneous palpation at the radial artery and auscultation using a stethoscope at the PMI, near the heart apex, for example. Typically, in people with pulse deficit, heart beats do not result in pulsations at the periphery, meaning the pulse rate is lower than the heart rate. Pulse deficit has been found to be significant in the context of premature ventricular contraction[citation needed] and atrial fibrillation.[8]

Rhythm

edit

A normal pulse is regular in rhythm and force. An irregular pulse may be due to sinus arrhythmia, ectopic beats, atrial fibrillation, paroxysmal atrial tachycardia, atrial flutter, partial heart block etc. Intermittent dropping out of beats at pulse is called "intermittent pulse". Examples of regular intermittent (regularly irregular) pulse include pulsus bigeminus, second-degree atrioventricular block. An example of irregular intermittent (irregularly irregular) pulse is atrial fibrillation.

Volume

edit

The degree of expansion displayed by artery during diastolic and systolic state is called volume. It is also known as amplitude, expansion or size of pulse.

Hypokinetic pulse

edit

A weak pulse signifies narrow pulse pressure. It may be due to low cardiac output (as seen in shock, congestive cardiac failure), hypovolemia, valvular heart disease (such as aortic outflow tract obstruction, mitral stenosis, aortic arch syndrome) etc.

Hyperkinetic pulse

edit

A bounding pulse signifies high pulse pressure. It may be due to low peripheral resistance (as seen in fever, anemia, thyrotoxicosis, hyperkinetic heart syndrome [de], A-V fistula, Paget's disease, beriberi, liver cirrhosis), increased cardiac output, increased stroke volume (as seen in anxiety, exercise, complete heart block, aortic regurgitation), decreased distensibility of arterial system (as seen in atherosclerosis, hypertension and coarctation of aorta).

The strength of the pulse can also be reported:[9][10]

  • 0 = Absent
  • 1 = Barely palpable
  • 2 = Easily palpable
  • 3 = Full
  • 4 = Aneurysmal or bounding pulse

Force

edit

Also known as compressibility of pulse. It is a rough measure of systolic blood pressure.

Tension

edit

It corresponds to diastolic blood pressure. A low tension pulse (pulsus mollis), the vessel is soft or impalpable between beats. In high tension pulse (pulsus durus), vessels feel rigid even between pulse beats.

Form

edit

A form or contour of a pulse is palpatory estimation of arteriogram. A quickly rising and quickly falling pulse (pulsus celer) is seen in aortic regurgitation. A slow rising and slowly falling pulse (pulsus tardus) is seen in aortic stenosis.

Equality

edit

Comparing pulses and different places gives valuable clinical information.

A discrepant or unequal pulse between left and right radial artery is observed in anomalous or aberrant course of artery, coarctation of aorta, aortitis, dissecting aneurysm, peripheral embolism etc. An unequal pulse between upper and lower extremities is seen in coarctation to aorta, aortitis, block at bifurcation of aorta, dissection of aorta, iatrogenic trauma and arteriosclerotic obstruction.

Condition of arterial wall

edit

A normal artery is not palpable after flattening by digital pressure. A thick radial artery which is palpable 7.5–10 cm up the forearm is suggestive of arteriosclerosis.

Radio-femoral delay

edit

In coarctation of aorta, femoral pulse may be significantly delayed as compared to radial pulse (unless there is coexisting aortic regurgitation). The delay can also be observed in supravalvar aortic stenosis.

Patterns

edit

Several pulse patterns can be of clinical significance. These include:

  • Anacrotic pulse: notch on the upstroke of the carotid pulse. Two distinct waves (slow initial upstroke and delayed peak, which is close to S2). Present in AS.
  • Dicrotic pulse: is characterized by two beats per cardiac cycle, one systolic and the other diastolic. Physiologically, the dicrotic wave is the result of reflected waves from the lower extremities and aorta. Conditions associated with low cardiac output and high systemic vascular resistance can produce a dicrotic pulse.[11][12]
  • Pulse deficit: difference in the heart rate by direct cardiac ausculation and by palpation of the peripheral arterial pulse rate when in atrial fibrillation (AF).
  • Pulsus alternans: an ominous medical sign that indicates progressive systolic heart failure. To trained fingertips, the examiner notes a pattern of a strong pulse followed by a weak pulse over and over again. This pulse signals a flagging effort of the heart to sustain itself in systole. It also can be detected in HCM with obstruction.
  • Pulsus bigeminus: indicates a pair of hoofbeats within each heartbeat. Concurrent auscultation of the heart may reveal a gallop rhythm of the native heartbeat.
  • Pulsus bisferiens: is characterized by two beats per cardiac cycle, both systolic, unlike the dicrotic pulse. It is an unusual physical finding typically seen in patients with aortic valve diseases if the aortic valve does not normally open and close. Trained fingertips will observe two pulses to each heartbeat instead of one.
  • Pulsus tardus et parvus, also pulsus parvus et tardus, slow-rising pulse and anacrotic pulse, is weak (parvus), and late (tardus) relative to its expected characteristics. It is caused by a stiffened aortic valve that makes it progressively harder to open, thus requiring increased generation of blood pressure in the left ventricle. It is seen in aortic valve stenosis.[12][13][14][15]
  • Pulsus paradoxus: a condition in which some heartbeats cannot be detected at the radial artery during the inspiration phase of respiration. It is caused by an exaggerated decrease in blood pressure during this phase, and is diagnostic of a variety of cardiac and respiratory conditions of varying urgency, such as cardiac tamponade.
  • Tachycardia: an elevated resting heart rate. In general an electrocardiogram (ECG) is required to identify the type of tachycardia.
  • Pulsatile This description of the pulse implies the intrinsic physiology of systole and diastole. Scientifically, systole and diastole are forces that expand and contract the pulmonary and systemic circulations.
  • A collapsing pulse is a sign of hyperdynamic circulation, which can be seen in AR or PDA.

Common palpable sites

edit

Sites can be divided into peripheral pulses and central pulses. Central pulses include the carotid, femoral, and brachial pulses.[2]

Upper limb

edit
 
Front of right upper extremity
  • Axillary pulse: located inferiorly of the lateral wall of the axilla
  • Brachial pulse: located on the inside of the upper arm near the elbow, frequently used in place of carotid pulse in infants (brachial artery)
  • Radial pulse: located on the lateral of the wrist (radial artery). It can also be found in the anatomical snuff box. Commonly, the radial pulse is measured with three fingers. The finger closest to the heart is used to occlude the pulse pressure, the middle finger is used get a crude estimate of the blood pressure, and the finger most distal to the heart (usually the ring finger) is used to nullify the effect of the ulnar pulse as the two arteries are connected via the palmar arches (superficial and deep).[citation needed]
  • Ulnar pulse: located on the medial of the wrist (ulnar artery).

Lower limb

edit
  • Femoral pulse: located in the inner thigh, at the mid-inguinal point, halfway between the pubic symphysis and anterior superior iliac spine (femoral artery).
  • Popliteal pulse: Above the knee in the popliteal fossa, found by holding the bent knee. The patient bends the knee at approximately 124°, and the health care provider holds it in both hands to find the popliteal artery in the pit behind the knee (popliteal artery).
  • Dorsalis pedis pulse: located on top of the foot, immediately lateral to the extensor of hallucis longus (dorsalis pedis artery).
  • Tibialis posterior pulse: located on the medial side of the ankle, 2 cm inferior and 2 cm posterior to the medial malleolus (posterior tibial artery). It is easily palpable over Pimenta's Point.

Head and neck

edit
 
Arteries of the neck.
  • Carotid pulse: located in the neck (carotid artery). The carotid artery should be palpated gently and while the patient is sitting or lying down. Stimulating its baroreceptors with low palpitation can provoke severe bradycardia or even stop the heart in some sensitive persons. Also, a person's two carotid arteries should not be palpated at the same time. Doing so may limit the flow of blood to the head, possibly leading to fainting or brain ischemia. It can be felt between the anterior border of the sternocleidomastoid muscle, above the hyoid bone and lateral to the thyroid cartilage.
  • Facial pulse: located on the mandible (lower jawbone) on a line with the corners of the mouth (facial artery).
  • Temporal pulse: located on the temple directly in front of the ear (superficial temporal artery).

Although the pulse can be felt in multiple places in the head, people should not normally hear their heartbeats within the head. This is called pulsatile tinnitus, and it can indicate several medical disorders.

Torso

edit
  • Apical pulse: located in the 5th left intercostal space, 1.25 cm lateral to the mid-clavicular line. In contrast with other pulse sites, the apical pulse site is unilateral, and measured not under an artery, but below the heart itself (more specifically, the apex of the heart). See also apex beat.

History

edit

Pulse rate was first measured by ancient Greek physicians and scientists. The first person to measure the heart beat was Herophilus of Alexandria, Egypt (c. 335–280 BC) who designed a water clock to time the pulse.[16] Rumi has mentioned in a poem that "The wise physician measured the patient's pulse and became aware of his condition." It shows the practice was common during Rumi's era and geography.[17] The first person to accurately measure the pulse rate was Santorio Santorii who invented the pulsilogium, a form of pendulum which was later studied by Galileo Galilei.[18] A century later another physician, de Lacroix, used the pulsilogium to test cardiac function.

See also

edit

References

edit
  1. ^ a b "pulse". The Free Dictionary.
  2. ^ a b Berg, Dale; Worzala, Katherine (2006). Atlas of Adult Physical Diagnosis. Lippincott Williams & Wilkins. p. 80. ISBN 9780781741903. Archived from the original on 2023-07-03. Retrieved 2020-10-03.
  3. ^ "Pulse". MedlinePlus Medical Encyclopedia. Retrieved 25 November 2024.
  4. ^ "sphygmograph". The Free Dictionary.
  5. ^ Temkin 165;BBC[a]
  6. ^ a b Yuda, Emi; Yamamoto, Kento; Yoshida, Yutaka; Hayano, Junichiro (21 February 2020). "Differences in pulse rate variability with measurement site". Journal of Physiological Anthropology. 39 (1): 4. doi:10.1186/s40101-020-0214-1. ISSN 1880-6791. PMC 7035641. PMID 32085811.
  7. ^ "Pulse Rate Measurement". Healthwise. WebMD. Archived from the original on 23 July 2012. Retrieved 14 March 2011.
  8. ^ Karadavut, Serhat; Altintop, Ismail (30 November 2021). "Pulse deficit in atrial fibrillation — a different perspective on rhythm or rate control strategy". Kardiologia Polska. 79 (11): 1231–1238. doi:10.33963/KP.a2021.0107. PMID 34506629.
  9. ^ "www.meddean.luc.edu". Archived from the original on 2008-09-07. Retrieved 2009-05-20.
  10. ^ "Vascular Surgery, University of Kansas School of Medicine". Archived from the original on 2009-02-04. Retrieved 2009-05-20.
  11. ^ Dennis, Mark; Bowen, William Talbot; Cho, Lucy (2016). Mechanisms of Clinical Signs – EPub3. Elsevier Health Sciences. p. 177. ISBN 9780729585613. Archived from the original on 2023-07-03. Retrieved 2018-01-02.
  12. ^ a b McGee, Steven (2016). Evidence-Based Physical Diagnosis E-Book. Elsevier Health Sciences. pp. 105–106. ISBN 9780323508711. Archived from the original on 2023-07-03. Retrieved 2018-01-02.
  13. ^ Li JC, Yuan Y, Qin W, et al. (April 2007). "Evaluation of the tardus-parvus pattern in patients with atherosclerotic and nonatherosclerotic renal artery stenosis". J Ultrasound Med. 26 (4): 419–26. doi:10.7863/jum.2007.26.4.419. PMID 17384038. S2CID 11562194. Archived from the original on 2020-05-10. Retrieved 2018-01-02.
  14. ^ Toy, Eugene, et al. Case Files: Internal Medicine. McGraw-Hill Companies, Inc. 2007. Page 43. ISBN 0-07-146303-8.
  15. ^ Sanders, Roger C.; Winter, Thomas Charles (2007). Clinical Sonography: A Practical Guide. Lippincott Williams & Wilkins. p. 219. ISBN 9780781748698. Archived from the original on 2023-07-03. Retrieved 2018-01-02.
  16. ^ Hajar, R. (2018). "The Pulse from Ancient to Modern Medicine". Heart Views. 19 (3): 117–120. doi:10.4103/HEARTVIEWS.HEARTVIEWS_16_19. PMC 6448473. PMID 31007864.
  17. ^ "Rumi poem on the afflicted patient". Archived from the original on 2022-12-16. Retrieved 2022-12-16.
  18. ^ Bigotti, Fabrizio; Taylor, David (2017). "The Pulsilogium of Santorio: New Light on Technology and Measurement in Early Modern Medicine". Societate Si Politica. 11 (2): 53–113. ISSN 1843-1348. PMC 6407692. PMID 30854144.
edit
  NODES
HOME 1
inspiration 1
Intern 1
iOS 3
languages 1
Note 2
Observable 1
os 36
server 1
text 1
visual 2
web 1