Noncontact simultaneous dual wavelength photoplethysmography: a further step toward noncontact pulse oximetry
- PMID: 17477684
- DOI: 10.1063/1.2724789
Noncontact simultaneous dual wavelength photoplethysmography: a further step toward noncontact pulse oximetry
Abstract
We present a camera-based device capable of capturing two photoplethysmographic (PPG) signals at two different wavelengths simultaneously, in a remote noncontact manner. The system comprises a complementary metal-oxide semiconductor camera and dual wavelength array of light emitting diodes (760 and 880 nm). By alternately illuminating a region of tissue with each wavelength of light, and detecting the backscattered photons with the camera at a rate of 16 frames/wavelength s, two multiplexed PPG wave forms are simultaneously captured. This process is the basis of pulse oximetry, and we describe how, with the inclusion of a calibration procedure, this system could be used as a noncontact pulse oximeter to measure arterial oxygen saturation (S(p)O(2)) remotely. Results from an experiment on ten subjects, exhibiting normal S(p)O(2) readings, that demonstrate the instrument's ability to capture signals from a range of subjects under realistic lighting and environmental conditions are presented. We compare the signals captured by the noncontact system to a conventional PPG signal captured concurrently from a finger, and show by means of a J. Bland and D. Altman [Lancet 327, 307 (1986); Statistician 32, 307 (1983)] test, the noncontact device to be comparable to a contact device as a monitor of heart rate. We highlight some considerations that should be made when using camera-based "integrative" sampling methods and demonstrate through simulation, the suitability of the captured PPG signals for application of existing pulse oximetry calibration procedures.
Similar articles
-
An in vivo investigation of photoplethysmographic signals and preliminary pulse oximetry estimation from the bowel using a new fiberoptic sensor.Anesth Analg. 2011 May;112(5):1104-9. doi: 10.1213/ANE.0b013e31820f8df3. Epub 2011 Feb 23. Anesth Analg. 2011. PMID: 21346164
-
Pulse oximetry in the oesophagus.Physiol Meas. 2006 Jan;27(1):R1-35. doi: 10.1088/0967-3334/27/1/R01. Epub 2005 Nov 29. Physiol Meas. 2006. PMID: 16365505 Review.
-
Contactless multiple wavelength photoplethysmographic imaging: a first step toward "SpO2 camera" technology.Ann Biomed Eng. 2005 Aug;33(8):1034-41. doi: 10.1007/s10439-005-5763-2. Ann Biomed Eng. 2005. PMID: 16133912
-
Variability in time delay between two models of pulse oximeters for deriving the photoplethysmographic signals.Physiol Meas. 2005 Aug;26(4):531-44. doi: 10.1088/0967-3334/26/4/017. Epub 2005 May 10. Physiol Meas. 2005. PMID: 15886446 Clinical Trial.
-
Pulse oximetry and photoplethysmographic waveform analysis of the esophagus and bowel.Curr Opin Anaesthesiol. 2008 Dec;21(6):779-83. doi: 10.1097/ACO.0b013e328317794d. Curr Opin Anaesthesiol. 2008. PMID: 18997529 Review.
Cited by
-
Ninety years of pulse oximetry: history, current status, and outlook.J Biomed Opt. 2024 Jun;29(Suppl 3):S33307. doi: 10.1117/1.JBO.29.S3.S33307. Epub 2024 Aug 17. J Biomed Opt. 2024. PMID: 39156662 Free PMC article. Review.
-
Effectiveness of consumer-grade contactless vital signs monitors: a systematic review and meta-analysis.J Clin Monit Comput. 2022 Feb;36(1):41-54. doi: 10.1007/s10877-021-00734-9. Epub 2021 Jul 9. J Clin Monit Comput. 2022. PMID: 34240262 Free PMC article. Review.
-
Diagnostic Features and Potential Applications of PPG Signal in Healthcare: A Systematic Review.Healthcare (Basel). 2022 Mar 16;10(3):547. doi: 10.3390/healthcare10030547. Healthcare (Basel). 2022. PMID: 35327025 Free PMC article. Review.
-
Reflection-type photoplethysmography pulse sensor based on an integrated optoelectronic chip with a ring structure.Biomed Opt Express. 2021 Sep 15;12(10):6277-6283. doi: 10.1364/BOE.437805. eCollection 2021 Oct 1. Biomed Opt Express. 2021. PMID: 34745736 Free PMC article.
-
Remote measurements of heart and respiration rates for telemedicine.PLoS One. 2013 Oct 8;8(10):e71384. doi: 10.1371/journal.pone.0071384. eCollection 2013. PLoS One. 2013. PMID: 24115996 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources