Quantitative characterization of subject motion in HR-pQCT images of the distal radius and tibia
- PMID: 21421091
- PMCID: PMC3108045
- DOI: 10.1016/j.bone.2011.03.755
Quantitative characterization of subject motion in HR-pQCT images of the distal radius and tibia
Abstract
Image quality degradation due to subject motion is a common artifact affecting in vivo high-resolution peripheral quantitative computed tomography (HR-pQCT) of bones. These artifacts confound the accuracy and reproducibility of bone density, geometry, and cortical and trabecular structure measurements. Observer-based systems for grading image quality and criteria for deciding when to repeat an acquisition and post hoc data quality control remain highly subjective and non-standardized. This study proposes an objective, quantitative technique for measuring subject motion in HR-pQCT acquisitions from raw projection data, using image similarity measures applied to parallelized projections at 0° and 180°. A total of 88 HR-pQCT exams with repeated acquisitions of the distal radius (N = 54) or distal tibia (N = 34) of 49 women (age = 59 ± 14 year) and 3 men (46 ± 2 year) were retrospectively evaluated. All images were graded from 1 (no visible motion artifacts) to 5 (severe motion artifacts) according to the manufacturer-suggested image quality grading system. In addition, to serve as the reference case without motion artifacts, two cadaveric wrist and two ankle specimens were imaged twice with repositioning. The motion-induced error was calculated as the percent difference in each bone parameter for the paired scans with and without visually apparent motion artifacts. Quantitative motion estimates (QMEs) for each motion-degraded scan were calculated using two different image similarity measures: sum of squared differences (SSD) and normalized cross-correlation (NCC). The mean values of QME(SSD) and QME(NCC) increased with the image quality grade for both radius and tibia. Quality grades were differentiated between grades 2 and 3 using QME(SSD), but not with QME(NCC), in addition to between grades 4 and 5. Both QMEs correlated significantly to the motion-induced errors in the measurements and their empirical relationship was derived. Subject motion had greater impact on the precision of trabecular structure indices than on the densitometric indices. The results of this study may provide a basis for establishing a threshold for motion artifacts in accordance to the study design as well as a standardized quality control protocol across operators and imaging centers.
Copyright © 2011 Elsevier Inc. All rights reserved.
Figures
Similar articles
-
Short-term in vivo precision of BMD and parameters of trabecular architecture at the distal forearm and tibia.Osteoporos Int. 2012 Aug;23(8):2151-8. doi: 10.1007/s00198-011-1829-1. Epub 2011 Dec 6. Osteoporos Int. 2012. PMID: 22143491 Clinical Trial.
-
Visual grading of motion induced image degradation in high resolution peripheral computed tomography: impact of image quality on measures of bone density and micro-architecture.Bone. 2012 Jan;50(1):111-8. doi: 10.1016/j.bone.2011.10.003. Epub 2011 Oct 13. Bone. 2012. PMID: 22019605
-
In vivo precision of three HR-pQCT-derived finite element models of the distal radius and tibia in postmenopausal women.BMC Musculoskelet Disord. 2016 Sep 13;17(1):389. doi: 10.1186/s12891-016-1238-x. BMC Musculoskelet Disord. 2016. PMID: 27619649 Free PMC article.
-
Recommendations for High-resolution Peripheral Quantitative Computed Tomography Assessment of Bone Density, Microarchitecture, and Strength in Pediatric Populations.Curr Osteoporos Rep. 2023 Oct;21(5):609-623. doi: 10.1007/s11914-023-00811-9. Epub 2023 Jul 10. Curr Osteoporos Rep. 2023. PMID: 37428435 Free PMC article. Review.
-
A systematic review and meta-analysis of pediatric normative peripheral quantitative computed tomography data.Bone Rep. 2021 Jul 7;15:101103. doi: 10.1016/j.bonr.2021.101103. eCollection 2021 Dec. Bone Rep. 2021. PMID: 34377749 Free PMC article. Review.
Cited by
-
High-resolution peripheral quantitative computed tomography for the assessment of bone strength and structure: a review by the Canadian Bone Strength Working Group.Curr Osteoporos Rep. 2013 Jun;11(2):136-46. doi: 10.1007/s11914-013-0140-9. Curr Osteoporos Rep. 2013. PMID: 23525967 Free PMC article. Review.
-
A customized protocol to assess bone quality in the metacarpal head, metacarpal shaft and distal radius: a high resolution peripheral quantitative computed tomography precision study.BMC Musculoskelet Disord. 2013 Dec 24;14:367. doi: 10.1186/1471-2474-14-367. BMC Musculoskelet Disord. 2013. PMID: 24364867 Free PMC article.
-
Deficits in Bone Geometry in Growth Hormone-Deficient Prepubertal Boys Revealed by High-Resolution Peripheral Quantitative Computed Tomography.Horm Res Paediatr. 2019;92(5):293-301. doi: 10.1159/000506229. Epub 2020 Mar 30. Horm Res Paediatr. 2019. PMID: 32224610 Free PMC article.
-
Short-term in vivo precision of BMD and parameters of trabecular architecture at the distal forearm and tibia.Osteoporos Int. 2012 Aug;23(8):2151-8. doi: 10.1007/s00198-011-1829-1. Epub 2011 Dec 6. Osteoporos Int. 2012. PMID: 22143491 Clinical Trial.
-
[Review of high-resolution peripheral quantitative computed tomography for the assessment of bone microstructure and strength].Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2018 Jun 25;35(3):468-474. doi: 10.7507/1001-5515.201707068. Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2018. PMID: 29938957 Free PMC article. Review. Chinese.
References
-
- MacNeil JA, Boyd SK. Accuracy of high-resolution peripheral quantitative computed tomography for measurement of bone quality. Med Eng Phys. 2007;29:1096–105. - PubMed
-
- MacNeil JA, Boyd SK. Improved reproducibility of high-resolution peripheral quantitative computed tomography for measurement of bone quality. Med Eng Phys. 2008;30:792–9. - PubMed