Semi-automatic assessment of skin capillary density: proof of principle and validation
- PMID: 23988877
- DOI: 10.1016/j.mvr.2013.08.003
Semi-automatic assessment of skin capillary density: proof of principle and validation
Abstract
Background: Skin capillary density and recruitment have been proven to be relevant measures of microvascular function. Unfortunately, the assessment of skin capillary density from movie files is very time-consuming, since this is done manually. This impedes the use of this technique in large-scale studies. We aimed to develop a (semi-) automated assessment of skin capillary density.
Methods: CapiAna (Capillary Analysis) is a newly developed semi-automatic image analysis application. The technique involves four steps: 1) movement correction, 2) selection of the frame range and positioning of the region of interest (ROI), 3) automatic detection of capillaries, and 4) manual correction of detected capillaries. To gain insight into the performance of the technique, skin capillary density was measured in twenty participants (ten women; mean age 56.2 [42-72] years). To investigate the agreement between CapiAna and the classic manual counting procedure, we used weighted Deming regression and Bland-Altman analyses. In addition, intra- and inter-observer coefficients of variation (CVs), and differences in analysis time were assessed.
Results: We found a good agreement between CapiAna and the classic manual method, with a Pearson's correlation coefficient (r) of 0.95 (P<0.001) and a Deming regression coefficient of 1.01 (95%CI: 0.91; 1.10). In addition, we found no significant differences between the two methods, with an intercept of the Deming regression of 1.75 (-6.04; 9.54), while the Bland-Altman analysis showed a mean difference (bias) of 2.0 (-13.5; 18.4) capillaries/mm(2). The intra- and inter-observer CVs of CapiAna were 2.5% and 5.6% respectively, while for the classic manual counting procedure these were 3.2% and 7.2%, respectively. Finally, the analysis time for CapiAna ranged between 25 and 35min versus 80 and 95min for the manual counting procedure.
Conclusion: We have developed a semi-automatic image analysis application (CapiAna) for the assessment of skin capillary density, which agrees well with the classic manual counting procedure, is time-saving, and has a better reproducibility as compared to the classic manual counting procedure. As a result, the use of skin capillaroscopy is feasible in large-scale studies, which importantly extends the possibilities to perform microcirculation research in humans.
© 2013.
Comment in
-
Comment to article: Semi-automatic assessment of skin capillary density: proof of principle and validation.Microvasc Res. 2014 May;93:21-2. doi: 10.1016/j.mvr.2014.02.010. Epub 2014 Mar 1. Microvasc Res. 2014. PMID: 24594439 No abstract available.
-
Response to comment on: semi-automatic assessment of skin capillary density: proof of principle and validation.Microvasc Res. 2014 Jul;94:7-8. doi: 10.1016/j.mvr.2014.04.006. Epub 2014 Apr 29. Microvasc Res. 2014. PMID: 24788072 No abstract available.
Similar articles
-
Quantification of capillary blood cell flow using reflectance confocal microscopy.Skin Res Technol. 2014 Aug;20(3):373-8. doi: 10.1111/srt.12128. Epub 2014 Feb 8. Skin Res Technol. 2014. PMID: 24506277
-
Image enhancement effect on inter and intra-observer reliability of nailfold capillary assessment.Microvasc Res. 2018 Nov;120:100-110. doi: 10.1016/j.mvr.2018.06.005. Epub 2018 Jun 27. Microvasc Res. 2018. PMID: 29958863
-
Elliptical broken line method for calculating capillary density in nailfold capillaroscopy: Proposal and evaluation.Microvasc Res. 2017 Sep;113:1-8. doi: 10.1016/j.mvr.2017.04.002. Epub 2017 Apr 14. Microvasc Res. 2017. PMID: 28414021
-
New techniques for clinical assessment of the peripheral microcirculation.Drugs. 1999;59 Spec No:17-22. Drugs. 1999. PMID: 10548388 Review.
-
Visualization of dermal blood vessels--capillaroscopy.Clin Exp Dermatol. 1999 Nov;24(6):473-8. doi: 10.1046/j.1365-2230.1999.00537.x. Clin Exp Dermatol. 1999. PMID: 10606953 Review.
Cited by
-
Visualization of skin capillaries with moving red blood cells in arbitrary area of the body.Biomed Opt Express. 2019 Aug 30;10(9):4896-4906. doi: 10.1364/BOE.10.004896. eCollection 2019 Sep 1. Biomed Opt Express. 2019. PMID: 31565533 Free PMC article.
-
Vision transformer assisting rheumatologists in screening for capillaroscopy changes in systemic sclerosis: an artificial intelligence model.Rheumatology (Oxford). 2023 Jul 5;62(7):2492-2500. doi: 10.1093/rheumatology/keac541. Rheumatology (Oxford). 2023. PMID: 36347487 Free PMC article.
-
Observation and density estimation of a large number of skin capillaries using wide-field portable video capillaroscopy and semantic segmentation.J Biomed Opt. 2023 Oct;28(10):106003. doi: 10.1117/1.JBO.28.10.106003. Epub 2023 Oct 24. J Biomed Opt. 2023. PMID: 37881371 Free PMC article.
-
Skeletal Muscle, Skin, and Bone as Three Major Nitrate Reservoirs in Mammals: Chemiluminescence and 15N-Tracer Studies in Yorkshire Pigs.Nutrients. 2024 Aug 13;16(16):2674. doi: 10.3390/nu16162674. Nutrients. 2024. PMID: 39203815 Free PMC article.
-
A Rare Case of Intermittent Claudication Associated with Impaired Arterial Vasodilation.Case Rep Vasc Med. 2017;2017:4868123. doi: 10.1155/2017/4868123. Epub 2017 Dec 24. Case Rep Vasc Med. 2017. PMID: 29435381 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials