Accuracy and reproducibility of bending stiffness measurements by mechanical response tissue analysis in artificial human ulnas
- PMID: 25261885
- DOI: 10.1016/j.jbiomech.2014.09.005
Accuracy and reproducibility of bending stiffness measurements by mechanical response tissue analysis in artificial human ulnas
Abstract
Osteoporosis is characterized by reduced bone strength, but no FDA-approved medical device measures bone strength. Bone strength is strongly associated with bone stiffness, but no FDA-approved medical device measures bone stiffness either. Mechanical Response Tissue Analysis (MRTA) is a non-significant risk, non-invasive, radiation-free, vibration analysis technique for making immediate, direct functional measurements of the bending stiffness of long bones in humans in vivo. MRTA has been used for research purposes for more than 20 years, but little has been published about its accuracy. To begin to investigate its accuracy, we compared MRTA measurements of bending stiffness in 39 artificial human ulna bones to measurements made by Quasistatic Mechanical Testing (QMT). In the process, we also quantified the reproducibility (i.e., precision and repeatability) of both methods. MRTA precision (1.0±1.0%) and repeatability (3.1 ± 3.1%) were not as high as those of QMT (0.2 ± 0.2% and 1.3+1.7%, respectively; both p<10(-4)). The relationship between MRTA and QMT measurements of ulna bending stiffness was indistinguishable from the identity line (p=0.44) and paired measurements by the two methods agreed within a 95% confidence interval of ± 5%. If such accuracy can be achieved on real human ulnas in situ, and if the ulna is representative of the appendicular skeleton, MRTA may prove clinically useful.
Keywords: Bending stiffness; Composite ulna; Mechanical response tissue analysis; Quasistatic mechanical testing.
Copyright © 2014 Elsevier Ltd. All rights reserved.
Similar articles
-
Response to "Clinical Evaluation of Bone Strength and Fracture Risk".Curr Osteoporos Rep. 2017 Aug;15(4):396-397. doi: 10.1007/s11914-017-0386-8. Curr Osteoporos Rep. 2017. PMID: 28660374 Free PMC article. Review.
-
A new noninvasive mechanical bending test accurately predicts ulna bending strength in cadaveric human arms.Bone. 2019 Mar;120:336-346. doi: 10.1016/j.bone.2018.11.018. Epub 2018 Nov 26. Bone. 2019. PMID: 30496886
-
Noninvasive determination of bone mechanical properties using vibration response: a refined model and validation in vivo.J Biomech. 1996 Jan;29(1):91-8. doi: 10.1016/0021-9290(95)00030-5. J Biomech. 1996. PMID: 8839021
-
Noninvasive determination of ulnar stiffness from mechanical response--in vivo comparison of stiffness and bone mineral content in humans.J Biomech Eng. 1988 May;110(2):87-96. doi: 10.1115/1.3108423. J Biomech Eng. 1988. PMID: 3379938
-
Rationale, design and clinical performance of the mechanical response tissue analyser: a non-invasive technology for measurement of long bone bending stiffness.J Med Eng Technol. 2013 Feb;37(2):144-9. doi: 10.3109/03091902.2012.753128. J Med Eng Technol. 2013. PMID: 23360197 Review.
Cited by
-
A novel method for bone fatigue monitoring and prediction.Bone Rep. 2019 Aug 17;11:100221. doi: 10.1016/j.bonr.2019.100221. eCollection 2019 Dec. Bone Rep. 2019. PMID: 31508458 Free PMC article.
-
Response to Loucks et al.'s Comment on "Clinical Evaluation of Bone Strength and Fracture Risk".Curr Osteoporos Rep. 2017 Aug;15(4):398. doi: 10.1007/s11914-017-0387-7. Curr Osteoporos Rep. 2017. PMID: 28733953 No abstract available.
-
In Vivo Assessment of Cortical Bone Fragility.Curr Osteoporos Rep. 2020 Feb;18(1):13-22. doi: 10.1007/s11914-020-00558-7. Curr Osteoporos Rep. 2020. PMID: 32088857 Free PMC article. Review.
-
Response to "Clinical Evaluation of Bone Strength and Fracture Risk".Curr Osteoporos Rep. 2017 Aug;15(4):396-397. doi: 10.1007/s11914-017-0386-8. Curr Osteoporos Rep. 2017. PMID: 28660374 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical