Cyclic Fatigue Resistance of Nickel-Titanium Reciprocating Instruments after Simulated Clinical Use
- PMID: 32818567
- DOI: 10.1016/j.joen.2020.08.010
Cyclic Fatigue Resistance of Nickel-Titanium Reciprocating Instruments after Simulated Clinical Use
Abstract
Introduction: The aim of this study was to evaluate the cyclic fatigue resistance of different heat-treated nickel-titanium reciprocating instruments in 2 different situations: new and used instruments after preparing 3 curved canals.
Methods: A total sample of 60 nickel-titanium instruments of 3 systems (n = 20 per system) were used in this study: ProDesign R (tip 25, 0.06 taper; Easy Dental Equipment, Belo Horizonte, Brazil), Reciproc Blue (tip 25, 0.08v taper; VDW, Munich, Germany), and WaveOne Gold (tip 25, 0.07v taper; Dentsply Sirona, Ballaigues, Switzerland). Thirty new instruments (n = 10 per system) were used to prepare 90 curved single-rooted mandibular premolars (n = 30). Each instrument was used to prepare 3 root canals, and after each canal preparation the instrument was ultrasonic cleaned and submitted to autoclave sterilization procedures. The other 30 instruments (n = 10 per system) were kept without use. Then, the new and used instruments were subjected to the cyclic fatigue test in an artificial canal with a 30° angle and a 5-mm radius of curvature. The time and number of cycles to fatigue were recorded. Data were analyzed using the unpaired t test for intragroup comparison. For intergroup comparison, analysis of variance and the Tukey test for multiple comparisons were used.
Results: The intergroup comparison of new instruments showed that ProDesign R had the highest cyclic fatigue resistance followed by Reciproc Blue and WaveOne Gold (P < .05). Regarding the used instruments, WaveOne Gold had the lowest cyclic fatigue resistance (P < .05). The intragroup comparison between new and used instruments showed that WaveOne Gold and ProDesign R presented a significant reduction in the cyclic fatigue resistance after simulated clinical use (P < .05); no difference was found with the Reciproc Blue instruments (P > .05).
Conclusions: ProDesign R had the highest cyclic fatigue resistance, whereas WaveOne Gold had the lowest for new and used instruments. Simulated clinical use affected the cyclic fatigue resistance of ProDesign R and WaveOne Gold but not of Reciproc Blue instruments.
Keywords: Cyclic fatigue; nickel-titanium alloy; reciprocating kinematics.
Copyright © 2020 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Cyclic fatigue using severely curved canals and torsional resistance of thermally treated reciprocating instruments.Clin Oral Investig. 2018 Sep;22(7):2633-2638. doi: 10.1007/s00784-018-2362-9. Epub 2018 Feb 1. Clin Oral Investig. 2018. PMID: 29392414
-
Comparison of Cyclic Fatigue Resistance of 5 Heat-treated Nickel-titanium Reciprocating Systems in Canals with Single and Double Curvatures.J Endod. 2019 Oct;45(10):1237-1241. doi: 10.1016/j.joen.2019.06.011. Epub 2019 Aug 13. J Endod. 2019. PMID: 31420109
-
Cyclic fatigue and torsional strength of three different thermally treated reciprocating nickel-titanium instruments.Clin Oral Investig. 2018 May;22(4):1865-1871. doi: 10.1007/s00784-017-2295-8. Epub 2017 Dec 9. Clin Oral Investig. 2018. PMID: 29224061
-
Influence of sodium hypochlorite on cyclic fatigue resistance of nickel-titanium instruments: A systematic review and meta-analysis of in vitro studies.Clin Oral Investig. 2023 Nov;27(11):6291-6319. doi: 10.1007/s00784-023-05243-4. Epub 2023 Sep 14. Clin Oral Investig. 2023. PMID: 37704917 Review.
-
Effect of body temperature on the cyclic fatigue resistance of the nickel-titanium endodontic instruments: A systematic review and meta-analysis of in vitro studies.J Conserv Dent. 2022 Jul-Aug;25(4):338-346. doi: 10.4103/jcd.jcd_55_22. Epub 2022 Aug 2. J Conserv Dent. 2022. PMID: 36187856 Free PMC article. Review.
Cited by
-
Cyclic Fatigue Resistance of Nickel-Titanium Rotary Instruments after Simulated Clinical Use.Appl Bionics Biomech. 2022 Oct 11;2022:1716008. doi: 10.1155/2022/1716008. eCollection 2022. Appl Bionics Biomech. 2022. PMID: 36267671 Free PMC article.
-
Effect of simulated clinical use and sterilization on the cyclic fatigue resistance of nickel titanium files.PeerJ. 2024 May 23;12:e17418. doi: 10.7717/peerj.17418. eCollection 2024. PeerJ. 2024. PMID: 38799069 Free PMC article.
-
Evaluation of the rotary kinematics between actual and set speeds of X-Smart Plus, VDW.Silver and iRoot motors.Braz Dent J. 2023 Mar-Apr;34(2):21-26. doi: 10.1590/0103-6440202304953. Braz Dent J. 2023. PMID: 37194853 Free PMC article.
-
Root Canal Instrumentation: Current Trends and Future Perspectives.Cureus. 2024 Apr 11;16(4):e58045. doi: 10.7759/cureus.58045. eCollection 2024 Apr. Cureus. 2024. PMID: 38738101 Free PMC article. Review.
-
Effect of number of uses and sterilization on the instrumented area and resistance of reciprocating instruments.Restor Dent Endod. 2021 Apr 29;46(2):e28. doi: 10.5395/rde.2021.46.e28. eCollection 2021 May. Restor Dent Endod. 2021. PMID: 34123764 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources